Breast ultrasound classifier β€” DenseNet121 v2b (research only)

⚠️ Not a medical device. Research model trained on public datasets; must not be used for diagnosis.

Five DenseNet121 fold models (v2b: mask-guided training + aspect-ratio crop augmentation) for benign vs malignant classification of breast ultrasound lesions. Code, data audit and full evaluation: https://github.com/wajidali99/breast-ultrasound-classification

Decision rule

Each model outputs P(malignant). A model votes malignant if P β‰₯ 0.335 (threshold frozen on BUSI out-of-fold predictions for β‰₯90% sensitivity). The label is the majority of the 5 votes.

Performance (ensemble AUC)

Data AUC
BUSI 5-fold CV (pooled out-of-fold) 0.940
BUSI locked test set (105 images) 0.970
BUS-BRA external (1,875 images, Brazil) 0.776 (95% CI 0.749–0.804)

Limitations

  • Large performance drop on an external hospital (AUC 0.97 β†’ 0.78): results do not transfer reliably across sites.

  • At the BUSI threshold, external sensitivity was 0.70 (target 0.90); site-specific calibration would be required.

  • Trained on ~530 BUSI images from a single centre; images without a lesion (normal) are out of scope.

  • The uploaded weights are a deterministic rebuild of the evaluated v2b models; four folds reproduce the original validation AUC exactly and fold 4 differs by 0.002. Reported BUS-BRA results come from the original v2 run.