DHEIVER commited on
Commit
5353470
1 Parent(s): d65d252

Upload 10 files

Browse files
README (7).md ADDED
@@ -0,0 +1,548 @@
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
1
+ ---
2
+ license: apache-2.0
3
+ language:
4
+ - en
5
+ pipeline_tag: summarization
6
+ widget:
7
+ - text: >-
8
+ the need for magnetic resonance imaging ( mri ) in patients with an
9
+ implanted pacemaker or implantable cardioverter - defibrillator ( icd ) is a
10
+ growing clinical issue . it is estimated that as many as 75% of active
11
+ cardiac device recipients will become indicated for mri . currently , the
12
+ vast majority of such devices are contraindicated for use with an mri . in
13
+ european heart rhythm association survey , published recently for non - mri
14
+ - certified icds ( 0.51.5 t field strength ) , the totally subcutaneous icd
15
+ ( s - icd ) system , an implantable defibrillator with no leads that touch
16
+ the heart , has recently been demonstrated to be a safe and effective
17
+ defibrillator option for patients at risk for sudden cardiac death . it
18
+ provides shock therapy and post - shock pacing therapy , but no long - term
19
+ bradycardia pacing . although it has been shown as an alternative to the
20
+ standard transvenous icd , its compatibility with mri remains unclear .
21
+ various types of clinical mri systems currently use a superconductive magnet
22
+ that creates a static magnetic field strength , typically 1.5 or 3 t. the
23
+ use of mri with most pacemakers and icds is considered a contraindication
24
+ due to potential hazards , including heating of the electrode that resides
25
+ in or on the heart , damage to myocardium , elevation of pacing thresholds ,
26
+ unintended induction of ventricular tachycardia ( vt ) or ventricular
27
+ fibrillation ( vf ) , pacing inhibition , permanent device malfunction , and
28
+ distortion of the mri scan . recently , mr - conditional. mr - conditional
29
+ indicates a lack of known hazards in a specified mri environment with
30
+ specified conditions of use . due to the variety of mri scanners and
31
+ scanning protocols , it is not practical to test even a single device under
32
+ all conditions . hence , mr - conditional labelling dictates that the device
33
+ is safe for use under certain scanning conditions , as well as how the
34
+ cardiac device should be programmed before an exposure to the magnetic field
35
+ in a mri scanner . the literature , although limited , provides some
36
+ guidance for imaging patients with implanted pacemakers or icds that do not
37
+ have mr - conditional labelling . this single - centre prospective non -
38
+ controlled study describes the first use of mri in patients with an
39
+ implanted s - icd . patients with implanted s - icd systems ( boston
40
+ scientific sqrx model 1010 and q - trak model 3010 ) were enrolled for mri
41
+ testing over a period of 18 months . the s - icd system implanted in this
42
+ patient cohort was composed of a can implanted in a left mid - lateral
43
+ pocket and a para - sternal subcutaneous electrode . the s - icd is
44
+ currently not certified for use with an mri ; therefore , the ethics
45
+ committee of homolka hospital , prague , czech republic approved our
46
+ clinical study . patients with newly implanted s - icd systems ( < 6 weeks )
47
+ were excluded , and none of the patients had any intravascular leads . the
48
+ patients were randomized for either a cardiac , brain , cervical , or lumbar
49
+ spinal scan . one of the subjects underwent an additional knee examination ,
50
+ due to reported chronic pain . a total of 15 patients were enrolled into
51
+ this study ( 12 males and three females , aged 2283 years , mean 53 years .
52
+ subjects in our cohort ( table 1 ) underwent a total of 22 mri scans between
53
+ 6 june 2012 and 24 december 2013 . in total , five brain scans , three
54
+ cardiac scans , 12 lumbar scans , one knee , and one cervical spine scan
55
+ were conducted ( table 2 ) . however , in one patient a minor disc
56
+ protrusion was found , in other mri revealed stenosis of intervertebral
57
+ foramen which was causing radicular pain of the nerve root l4 and based on
58
+ this examination the patient was referred to ct - navigated periradicular
59
+ therapy . table 1summary of patient anatomical data and scan locations ,
60
+ along with noted clinical eventsidagesexbmidgef , % indication for s -
61
+ icdheating0164f20.5hcmp / vfs85secondary preventionnone0283m30.0post - mi /
62
+ smvts post - catheter ablation/35secondary prevention ( post - transvenous
63
+ icd extraction)none0331m25.3arvc / d / smvts68secondary preventionin -
64
+ tolerable re - scanned0458m23.6post - mi / post - cabg30primary
65
+ preventionnone0577m25.5post - mi30primary preventionnone0663m27.0post -
66
+ mi30primary preventionnone0768m23.7post - mi / vfs / vts60secondary
67
+ prevention post - transvenous icd extraction / svc
68
+ occlusiontolerable0822m29.4brugada sy / vfs68secondary preventionin -
69
+ tolerable re - scanned0959m27.1dcmp / vfs / post - mitral valve
70
+ surgery/60secondary prev./post - transvenous icd extractionnone1041f24.6arvc
71
+ / d70primary preventionnone1123f21.5lqts / vf60secondary
72
+ preventionnone1266m36.9post - mi / vf / post - cabg50secondary prevention /
73
+ post - repeat transvenous icd extractiontolerable1348m22.9dcmp(non -
74
+ compaction)/vfs35secondary preventionnone1470m29systolic dysfunction of
75
+ lv35primary preventionnone1526m33brugada sy65primary preventionnonehcmp ,
76
+ hypertrophic cardiomyopathy ; smvt , sustained monomorphic ventricular
77
+ tachycardia ; mi , myocardial infarction ; arvc , arrhythmogenic right
78
+ ventricular cardiomyopathy ; cabg , coronary artery by - pass graft ; lqts ,
79
+ long qt syndrom . table 2parmeters of s - icd and patient sensation during
80
+ individual mri scansscan # idbody partheating sensationsshock zone (
81
+ b.p.m.)condit . shock zone ( b.p.m.)bat % episode
82
+ num.101brainnone2302101001202brainnone240220861303l spinein -
83
+ tolerable240220831403brainnone240220831504brainnone220190691605l
84
+ spinenone220210541706l spinenone240220681807l spinetolerable240220582908l
85
+ spinein - tolerablenananana1008brainnonenananana1108l
86
+ spinenone2302108411209heartnone2402208911310l
87
+ spinenone2301807911410heartnonenananana1511heartnone2301909711612l
88
+ spinetolerable2001709721712l spinenone2001709421813c
89
+ spinenone23019010041913l spinenone23019010042014l
90
+ spinenone2301908612115kneenone25021010012215l spinenone2502101001s - icd
91
+ parameters acquired prior- and post - mri were without any change ,
92
+ therefore only one value is presented.indices : na , not available ; l spine
93
+ , lumbar spine ; c spine , cervical spine . summary of patient anatomical
94
+ data and scan locations , along with noted clinical events hcmp ,
95
+ hypertrophic cardiomyopathy ; smvt , sustained monomorphic ventricular
96
+ tachycardia ; mi , myocardial infarction ; arvc , arrhythmogenic right
97
+ ventricular cardiomyopathy ; cabg , coronary artery by - pass graft ; lqts ,
98
+ long qt syndrom . parmeters of s - icd and patient sensation during
99
+ individual mri scans s - icd parameters acquired prior- and post - mri were
100
+ without any change , therefore only one value is presented . indices : na ,
101
+ not available ; l spine , lumbar spine ; c spine , cervical spine . studies
102
+ were performed using a siemens avanto 1.5 t mri scanner ( vb17 software ,
103
+ quantum gradient coils ) . all scans were run in normal operating mode ,
104
+ which is limited to 2 w / kg whole body averaged specific absorption rate (
105
+ sar ) . clinically relevant mri sequences were used for evaluation ( see
106
+ table 3 ) . table 3types of pulse sequences typically used for imaging of
107
+ respective anatomical areasscan locationscan
108
+ sequencesflairdwiflashfsehastesestirtruefispbrainxxxxheartxxxxcervical
109
+ spinexxkneexxxxlumbar spinexxflair , fluid attenuated inversion recovery ;
110
+ dwi , diffusion weighted imaging ; flash , fast low angle shot ; fse , fast
111
+ spin echo ; haste , half acquisition single - shot turbo spin echo ; se ,
112
+ spin echo ; stir , short tau inversion recovery ; truefisp , true fast
113
+ imaging with steady - state precession.fse sequence caused heating in
114
+ subjects with a thermistor probe during lumbar spine examination ( see the
115
+ text for details ) . types of pulse sequences typically used for imaging of
116
+ respective anatomical areas flair , fluid attenuated inversion recovery ;
117
+ dwi , diffusion weighted imaging ; flash , fast low angle shot ; fse , fast
118
+ spin echo ; haste , half acquisition single - shot turbo spin echo ; se ,
119
+ spin echo ; stir , short tau inversion recovery ; truefisp , true fast
120
+ imaging with steady - state precession . fse sequence caused heating in
121
+ subjects with a thermistor probe during lumbar spine examination ( see the
122
+ text for details ) . patients were asked to report immediately any pain ,
123
+ torqueing movement , or heating sensation in the area of the pocket or the
124
+ electrode by pressing an emergency bulb . furthermore , all patients were
125
+ questioned immediately following the mri procedure to ascertain any
126
+ discomfort in the vicinity of the can or electrode . pulse oximetry and
127
+ standard lead electrocardiogram ( ecg ) if discomfort occurred , the patient
128
+ was asked if the scan could be repeated at a later time using a revised scan
129
+ sequence or the subject was again randomized for another anatomical area .
130
+ since none of the components of the s - icd system are on or in the heart ,
131
+ heating near or around however , heating near the electrode or can with the
132
+ s - icd system may still cause serious patient discomfort . therefore ,
133
+ along with education of subjects , each patient was instrumented by taping
134
+ an oesophageal temperature probe ( beta - therm model g22k7mcd8 ) on the
135
+ skin over the mid - lateral implant site to record any temperature
136
+ excursions that might be correlated to patient symptoms of heating /
137
+ discomfort near the pocket . to minimize the risk of inappropriate therapy ,
138
+ the s - icd system was programmed to therapy each s - icd system was
139
+ evaluated prior to and immediately after the scan to verify proper
140
+ functioning , including interrogation , sensing , and battery voltage .
141
+ after the completion of the mri , long - term regular clinical follow - up
142
+ and checking of the device were performed . patients with implanted s - icd
143
+ systems ( boston scientific sqrx model 1010 and q - trak model 3010 ) were
144
+ enrolled for mri testing over a period of 18 months . the s - icd system
145
+ implanted in this patient cohort was composed of a can implanted in a left
146
+ mid - lateral pocket and a para - sternal subcutaneous electrode . the s -
147
+ icd is currently not certified for use with an mri ; therefore , the ethics
148
+ committee of homolka hospital , prague , czech republic approved our
149
+ clinical study . patients with newly implanted s - icd systems ( < 6 weeks )
150
+ were excluded , and none of the patients had any intravascular leads . the
151
+ patients were randomized for either a cardiac , brain , cervical , or lumbar
152
+ spinal scan . one of the subjects underwent an additional knee examination ,
153
+ due to reported chronic pain . a total of 15 patients were enrolled into
154
+ this study ( 12 males and three females , aged 2283 years , mean 53 years .
155
+ subjects in our cohort ( table 1 ) underwent a total of 22 mri scans between
156
+ 6 june 2012 and 24 december 2013 . in total , five brain scans , three
157
+ cardiac scans , 12 lumbar scans , one knee , and one cervical spine scan
158
+ were conducted ( table 2 ) . however , in one patient a minor disc
159
+ protrusion was found , in other mri revealed stenosis of intervertebral
160
+ foramen which was causing radicular pain of the nerve root l4 and based on
161
+ this examination the patient was referred to ct - navigated periradicular
162
+ therapy . table 1summary of patient anatomical data and scan locations ,
163
+ along with noted clinical eventsidagesexbmidgef , % indication for s -
164
+ icdheating0164f20.5hcmp / vfs85secondary preventionnone0283m30.0post - mi /
165
+ smvts post - catheter ablation/35secondary prevention ( post - transvenous
166
+ icd extraction)none0331m25.3arvc / d / smvts68secondary preventionin -
167
+ tolerable re - scanned0458m23.6post - mi / post - cabg30primary
168
+ preventionnone0577m25.5post - mi30primary preventionnone0663m27.0post -
169
+ mi30primary preventionnone0768m23.7post - mi / vfs / vts60secondary
170
+ prevention post - transvenous icd extraction / svc
171
+ occlusiontolerable0822m29.4brugada sy / vfs68secondary preventionin -
172
+ tolerable re - scanned0959m27.1dcmp / vfs / post - mitral valve
173
+ surgery/60secondary prev./post - transvenous icd extractionnone1041f24.6arvc
174
+ / d70primary preventionnone1123f21.5lqts / vf60secondary
175
+ preventionnone1266m36.9post - mi / vf / post - cabg50secondary prevention /
176
+ post - repeat transvenous icd extractiontolerable1348m22.9dcmp(non -
177
+ compaction)/vfs35secondary preventionnone1470m29systolic dysfunction of
178
+ lv35primary preventionnone1526m33brugada sy65primary preventionnonehcmp ,
179
+ hypertrophic cardiomyopathy ; smvt , sustained monomorphic ventricular
180
+ tachycardia ; mi , myocardial infarction ; arvc , arrhythmogenic right
181
+ ventricular cardiomyopathy ; cabg , coronary artery by - pass graft ; lqts ,
182
+ long qt syndrom . table 2parmeters of s - icd and patient sensation during
183
+ individual mri scansscan # idbody partheating sensationsshock zone (
184
+ b.p.m.)condit . shock zone ( b.p.m.)bat % episode
185
+ num.101brainnone2302101001202brainnone240220861303l spinein -
186
+ tolerable240220831403brainnone240220831504brainnone220190691605l
187
+ spinenone220210541706l spinenone240220681807l spinetolerable240220582908l
188
+ spinein - tolerablenananana1008brainnonenananana1108l
189
+ spinenone2302108411209heartnone2402208911310l
190
+ spinenone2301807911410heartnonenananana1511heartnone2301909711612l
191
+ spinetolerable2001709721712l spinenone2001709421813c
192
+ spinenone23019010041913l spinenone23019010042014l
193
+ spinenone2301908612115kneenone25021010012215l spinenone2502101001s - icd
194
+ parameters acquired prior- and post - mri were without any change ,
195
+ therefore only one value is presented.indices : na , not available ; l spine
196
+ , lumbar spine ; c spine , cervical spine . summary of patient anatomical
197
+ data and scan locations , along with noted clinical events hcmp ,
198
+ hypertrophic cardiomyopathy ; smvt , sustained monomorphic ventricular
199
+ tachycardia ; mi , myocardial infarction ; arvc , arrhythmogenic right
200
+ ventricular cardiomyopathy ; cabg , coronary artery by - pass graft ; lqts ,
201
+ long qt syndrom . parmeters of s - icd and patient sensation during
202
+ individual mri scans s - icd parameters acquired prior- and post - mri were
203
+ without any change , therefore only one value is presented . indices : na ,
204
+ not available ; l spine , lumbar spine ; c spine , cervical spine . studies
205
+ were performed using a siemens avanto 1.5 t mri scanner ( vb17 software ,
206
+ quantum gradient coils ) . all scans were run in normal operating mode ,
207
+ which is limited to 2 w / kg whole body averaged specific absorption rate (
208
+ sar ) . clinically relevant mri sequences were used for evaluation ( see
209
+ table 3 ) . table 3types of pulse sequences typically used for imaging of
210
+ respective anatomical areasscan locationscan
211
+ sequencesflairdwiflashfsehastesestirtruefispbrainxxxxheartxxxxcervical
212
+ spinexxkneexxxxlumbar spinexxflair , fluid attenuated inversion recovery ;
213
+ dwi , diffusion weighted imaging ; flash , fast low angle shot ; fse , fast
214
+ spin echo ; haste , half acquisition single - shot turbo spin echo ; se ,
215
+ spin echo ; stir , short tau inversion recovery ; truefisp , true fast
216
+ imaging with steady - state precession.fse sequence caused heating in
217
+ subjects with a thermistor probe during lumbar spine examination ( see the
218
+ text for details ) . types of pulse sequences typically used for imaging of
219
+ respective anatomical areas flair , fluid attenuated inversion recovery ;
220
+ dwi , diffusion weighted imaging ; flash , fast low angle shot ; fse , fast
221
+ spin echo ; haste , half acquisition single - shot turbo spin echo ; se ,
222
+ spin echo ; stir , short tau inversion recovery ; truefisp , true fast
223
+ imaging with steady - state precession . fse sequence caused heating in
224
+ subjects with a thermistor probe during lumbar spine examination ( see the
225
+ text for details ) . patients were asked to report immediately any pain ,
226
+ torqueing movement , or heating sensation in the area of the pocket or the
227
+ electrode by pressing an emergency bulb . furthermore , all patients were
228
+ questioned immediately following the mri procedure to ascertain any
229
+ discomfort in the vicinity of the can or electrode . pulse oximetry and
230
+ standard lead electrocardiogram ( ecg ) if discomfort occurred , the patient
231
+ was asked if the scan could be repeated at a later time using a revised scan
232
+ sequence or the subject was again randomized for another anatomical area .
233
+ since none of the components of the s - icd system are on or in the heart ,
234
+ heating near or around the electrode can not harm the myocardium . however ,
235
+ heating near the electrode or can with the s - icd system may still cause
236
+ serious patient discomfort . therefore , along with education of subjects ,
237
+ each patient was instrumented by taping an oesophageal temperature probe (
238
+ beta - therm model g22k7mcd8 ) on the skin over the mid - lateral implant
239
+ site to record any temperature excursions that might be correlated to
240
+ patient symptoms of heating / discomfort near the pocket . to minimize the
241
+ risk of inappropriate therapy , the s - icd system was programmed to therapy
242
+ each s - icd system was evaluated prior to and immediately after the scan to
243
+ verify proper functioning , including interrogation , sensing , and battery
244
+ voltage . after the completion of the mri , the s - icd system was
245
+ reprogrammed to original settings . long - term regular clinical follow - up
246
+ and checking of the device were performed . no anomalies were noted via
247
+ pulse oximetry or ecg during the scans for any of the patients . eleven of
248
+ 15 patients reported no sensation or pain from heating of the can , two of
249
+ 15 patients reported feeling some heating , and two patients reported
250
+ intolerable heating ( see table 2 ) . in patients with intolerable heating ,
251
+ the scan was halted within seconds and changed to a scan of the brain ,
252
+ which proceeded without incident . patient reports of heating in the
253
+ vicinity of the can occurred only during lumbar scans with a thermistor
254
+ probe ; no such reports occurred during scans of the brain , cardiac area ,
255
+ cervical spine , or without the probe . in two cases where heating in the
256
+ vicinity of the can was reported by the patient , the scan sequence was
257
+ altered to reduce the intensity of radiofrequency ( rf ) field exposure by
258
+ reducing the turbo factor ( e.g. from 21 to 7 ) , increasing the repetition
259
+ time ( e.g. to > 4000 ms ) , and reducing the flip angle ( e.g. from 170 to
260
+ 120 ) . the target values were chosen arbitrarily to maintain image contrast
261
+ ( flip angle ) and keep scan time at reasonable limits ( turbo factor and
262
+ repetition time ) . less heating was noted by patients after these
263
+ modifications to the scan parameters were made . 03 ) was observed to have a
264
+ skin lesion , appearing to be a circular rash or ulcer on the surface of the
265
+ skin over the can , approximately 35 mm in diameter . the cause of this skin
266
+ anomaly is not known ; it was later noted to have fully healed at a follow -
267
+ up 10 days after the scan . to ascertain the effect of heating due to the
268
+ instrumented thermistor catheter , the two patients who experienced the
269
+ heating ( examinations 9 and 16 , see table 2 ) were rescanned several weeks
270
+ later without the thermistor catheter in place ( examinations 11 and 17 ) .
271
+ first , modified sequence ( with even lower amount of energy deposited in
272
+ the tissue ) was used , which caused no heating . as no sensation was
273
+ reported by the subjects , they were asked to report even a minimal
274
+ discomfort , and the lumbar scans were performed using the same settings
275
+ that resulted in heating with the thermistor catheter in place in the first
276
+ imaging session . the results of the rescans revealed that no heating was
277
+ felt by the patients when the thermistor catheter was absent . there were no
278
+ noted changes to battery voltage , ability to detect the qrs signal or
279
+ stored diagnostic data . pacing thresholds can not be assessed by the s -
280
+ icd system , so this was not evaluated . none of the patients reported any
281
+ pulling or twisting of the can or pain from heating of the s - icd electrode
282
+ . for scans of the brain , lumbar spine , knee , and cervical spine , no
283
+ effect from image artefact was noted in the anatomical area of interest .
284
+ however , for scans of the cardiac area , image artefact was noted to
285
+ interfere with the ability to see parts of the left ventricle , though the
286
+ right ventricle of the heart was unaffected and could be imaged usefully .
287
+ this was due to the can and not the electrode ( see figure 1 ) ,
288
+ modifications to the protocol for the lumbar spine resulted in a lower
289
+ signal - to - noise ratio ; however , the images remain in diagnostic
290
+ quality ( see figure 2 ) . figure 1kinetic study in four - chamber view :
291
+ the systolic ( a and c ) and diastolic ( b and d ) images of cine sequences
292
+ , four - chamber view . the steady - state free precession ( ssfp ) sequence
293
+ ( a and b ) shows more artefacts . in ssfp kinetic study , an inflow of dark
294
+ blood from the left pulmonary veins was seen . it could be caused by s - icd
295
+ but also by metallic ring in mitral annulus . the spoiled gradient echo (
296
+ gre ) sequence ( c and d ) is better , but an artefact at the lateral wall
297
+ is obvious . figure 2lumbar spine imaging with icd : low sar t2 fse sequence
298
+ ( upper image ) compared with normal t2 fse in the same subject ( lower
299
+ image , for the scanning parameters see the discussion section ) . kinetic
300
+ study in four - chamber view : the systolic ( a and c ) and diastolic ( b
301
+ and d ) images of cine sequences , four - chamber view . the steady - state
302
+ free precession ( ssfp ) sequence ( a and b ) shows more artefacts . in ssfp
303
+ kinetic study , an inflow of dark blood from the left pulmonary veins was
304
+ seen . it could be caused by s - icd but also by metallic ring in mitral
305
+ annulus . the spoiled gradient echo ( gre ) sequence ( c and d ) is better ,
306
+ but an artefact at the lateral wall is obvious . lumbar spine imaging with
307
+ icd : low sar t2 fse sequence ( upper image ) compared with normal t2 fse in
308
+ the same subject ( lower image , for the scanning parameters see the
309
+ discussion section ) . there were no noted changes to battery voltage ,
310
+ ability to detect the qrs signal or stored diagnostic data . pacing
311
+ thresholds can not be assessed by the s - icd system , so this was not
312
+ evaluated . none of the patients reported any pulling or twisting of the can
313
+ or pain from heating of the s - icd electrode . for scans of the brain ,
314
+ lumbar spine , knee , and cervical spine , no effect from image artefact was
315
+ noted in the anatomical area of interest . however , for scans of the
316
+ cardiac area , image artefact was noted to interfere with the ability to see
317
+ parts of the left ventricle , though the right ventricle of the heart was
318
+ unaffected and could be imaged usefully . this was due to the can and not
319
+ the electrode ( see figure 1 ) , modifications to the protocol for the
320
+ lumbar spine resulted in a lower signal - to - noise ratio ; however , the
321
+ images remain in diagnostic quality ( see figure 2 ) . figure 1kinetic study
322
+ in four - chamber view : the systolic ( a and c ) and diastolic ( b and d )
323
+ images of cine sequences , four - chamber view . the steady - state free
324
+ precession ( ssfp ) sequence ( a and b ) shows more artefacts . in ssfp
325
+ kinetic study , an inflow of dark blood from the left pulmonary veins was
326
+ seen . it could be caused by s - icd but also by metallic ring in mitral
327
+ annulus . the spoiled gradient echo ( gre ) sequence ( c and d ) is better ,
328
+ but an artefact at the lateral wall is obvious . figure 2lumbar spine
329
+ imaging with icd : low sar t2 fse sequence ( upper image ) compared with
330
+ normal t2 fse in the same subject ( lower image , for the scanning
331
+ parameters see the discussion section ) . kinetic study in four - chamber
332
+ view : the systolic ( a and c ) and diastolic ( b and d ) images of cine
333
+ sequences , four - chamber view . the steady - state free precession ( ssfp
334
+ ) sequence ( a and b ) shows more artefacts . in ssfp kinetic study , an
335
+ inflow of dark blood from the left pulmonary veins was seen . it could be
336
+ caused by s - icd but also by metallic ring in mitral annulus . the spoiled
337
+ gradient echo ( gre ) sequence ( c and d ) is better , but an artefact at
338
+ the lateral wall is obvious . lumbar spine imaging with icd : low sar t2 fse
339
+ sequence ( upper image ) compared with normal t2 fse in the same subject (
340
+ lower image , there are several reports in the current literature about mr -
341
+ conditional pacemakers from several companies , but very limited reports
342
+ about mr - conditional icds . biotronik announced in late 2011 release of
343
+ their first mr - compatible icd device and defibrillator leads pro mri , but
344
+ in the conditions of use excluded scanning of the torso and focused more on
345
+ the extremities examination . in european heart rhythm association survey ,
346
+ 60% of centres did not implant any mri - certified icds , 34.3% implanted <
347
+ 10 icd devices , and only 5.6% implanted 10 and more icds ; one - fifth of
348
+ responders stated that mri - certified icds should be implanted in all
349
+ patients but lack of reimbursement was indicated as a possible obstacle to
350
+ implant more mri - certified pacemakers / icds by 47.1% of responding
351
+ centres . none of the components of the s - icd system are on or in the
352
+ heart . the s - icd depends less upon being in direct contact with the
353
+ myocardium to function and instead uses far - field sensing and stimulation
354
+ to provide the shock and post - shock pacing therapy . as a consequence ,
355
+ unlike transvenous systems heating near or around the electrode can not harm
356
+ the myocardium , which could present with possible safety consequences such
357
+ as an elevation in pacing thresholds or scarring of the myocardium , but it
358
+ may still cause serious patient discomfort . because the s - icd is larger
359
+ than modern transvenous icd 's , there may be more potential for the can to
360
+ experience heating due to the magnetic gradient or rf field . we report
361
+ results from what we believe is the first experience of mri scanning in
362
+ patients with an implanted s - icd and in various anatomical areas . overall
363
+ , mri was performed safely in all patients , which is in contrast to the
364
+ current literature with mri imaging in patients with electrical - active
365
+ devices which are not mri - conditional . in our study , the primary
366
+ clinically significant event attributable to the mri scan was the occurrence
367
+ of heating in the area of the pocket in the four patients that underwent
368
+ lumbar scans . it was not known if this was due to the s - icd can itself or
369
+ an artefact of the thermistor catheter used to measure skin temperature over
370
+ the pocket . this required a revision of our protocol , which was to re -
371
+ scan two of the patients who complained of heating . re - scanning of these
372
+ patients without the thermistor probe resulted in no complaints of heating ,
373
+ so it is assumed that the thermistor catheter itself heated during the
374
+ lumbar scans and caused the discomfort . as further evidence , all the
375
+ heating complaints occurred during rf - intensive scan sequences ( namely
376
+ fast spin echo ) with the temperature probe located axially near the centre
377
+ of the bore , where rf fields are the highest . the thermistor catheter is
378
+ constructed of insulated conductive cables connected to electrodes and
379
+ should couple to the rf fields efficiently , causing heating at the
380
+ electrodes and pain or damage on the surface of the skin where the probe was
381
+ placed over the s - icd can . if the heating was due to the s - icd can
382
+ itself , it would more likely occur during gradient - intensive scan
383
+ sequences ( which can generate eddy currents on can surfaces and internal
384
+ components ) and at locations in the bore where there are high gradient
385
+ fields , such as near the bore edges . however , when the patient was
386
+ scanned with gradient - intensive scan sequences ( e.g. flair dwi ) and with
387
+ the s - icd system in high gradient field locations in the bore ( e.g. such
388
+ as during a brain scan ) , patients did not detect any heating or discomfort
389
+ . in addition , the subcutaneous lead , which was not instrumented with a
390
+ thermistor catheter , never resulted in any heating sensation noted by the
391
+ patient , even when exactly the same sequence that resulted in heating in
392
+ the first session was used . the use of mri - compatible temperature
393
+ monitors such as fibre optic temperature probes would have provided better
394
+ confirmation of possible skin temperature elevation and would not have been
395
+ affected by the rf fields . for cardiac imaging , the main problem to solve
396
+ is metallic artefact , especially on the gradient - echo sequences . like in
397
+ research performed by nazarian et al . , several scan protocols were used to
398
+ see if any yielded different effects or reduced the qualitative extent of
399
+ artefact . gradient mode was changed from normal to whisper , resulting in
400
+ slower ramping of the field and therefore diminishing the changes of the
401
+ magnetic field in time . artefacts when present were limited to blurring of
402
+ the left ventricle during cardiac scans and most yielded clinically useful
403
+ information . standard interrogation of the s - icd revealed no adverse
404
+ effects upon the functioning of the system . while no adverse effects upon
405
+ the post - scan s - icd device function were noted , not all possible
406
+ scanning protocols were tested . it should be noted that , four of the s -
407
+ icd 's were exposed to repeat mri scans without adverse effects to device
408
+ function . in addition , because the s - icd does not provide long - term
409
+ bradycardia pacing , it is assumed that pacemaker - dependent patients would
410
+ not be implanted with this system . the inhibition of the pacemaker function
411
+ during the scanning sequence and possible pacing threshold changes are a
412
+ unique concern in patients implanted with transvenous icds . this study
413
+ included only 15 patients and 22 scans done on the same 1.5 t mri scanner .
414
+ thus , even these preliminary results should only be applied to 1.5 t mri
415
+ scanners ( similarly as reported in the present literature for other
416
+ implantable devices ) . device functionality was tested immediately after
417
+ the scan but not for long - term effects . in addition , not all device
418
+ functions were tested although the s - icd system does have a beeper /
419
+ interrogation warning if battery levels or memory irregularities occur .
420
+ however , patients were scheduled for regular check - up and no defect of
421
+ the device was observed in following 725 months ( mean observation time 18
422
+ months ) . delayed enhancement mri for determining cardiac scarring was also
423
+ not tested . also , there are other anatomical areas that were not evaluated
424
+ , such as shoulder and knees . this study included only 15 patients and 22
425
+ scans done on the same 1.5 t mri scanner . thus , even these preliminary
426
+ results should only be applied to 1.5 t mri scanners ( similarly as reported
427
+ in the present literature for other implantable devices ) . device
428
+ functionality was tested immediately after the scan but not for long - term
429
+ effects . in addition , not all device functions were tested although the s
430
+ - icd system does have a beeper / interrogation warning if battery levels or
431
+ memory irregularities occur . however , patients were scheduled for regular
432
+ check - up and no defect of the device was observed in following 725 months
433
+ ( mean observation time 18 months ) . delayed enhancement mri for
434
+ determining cardiac scarring was also not tested . also , there are other
435
+ anatomical areas that were not evaluated , such as shoulder and knees .
436
+ while more data are required to support a claim of mri - conditional , this
437
+ study is the study to demonstrate the feasibility of exposing s - icd
438
+ patients to mri using the scanning and monitor protocol described , with
439
+ some precautionary measures including : ( i ) programming the device therapy
440
+ off ; ( ii ) limiting the sar to 2.0 w / kg ; ( iii ) continuous monitoring
441
+ of the patients pulse oximetry and ecg by qualified personnel and especially
442
+ for any feelings of heating ; ( iv ) evaluate device function post scan ; (
443
+ v ) availability of full resuscitation facilities at the mri site . given
444
+ the variables of different mri scanners , the decision to perform mri on
445
+ patients with an implanted s - icd system should be balanced against the
446
+ potential risks . in our study , the only heating was very likely introduced
447
+ by not fully mri - compatible thermometer probe ; subjects rescanned without
448
+ the probe did not report any abnormalities during the scan of any body area
449
+ listed ( brain , cervical and lumbar spine , heart , and knee ) . this study
450
+ was supported by iga mz r nt12094/2011 , research project charles university
451
+ in prague , prvouk p34 and unce 204010/2012 . funding to pay the open access
452
+ publication charges for this article was provided by iga mz r nt12094/2011 .
453
+ example_title: Summarization Example 1
454
+ tags:
455
+ - medical
456
+ ---
457
+
458
+
459
+
460
+ # Model Card: T5 Large for Medical Text Summarization
461
+
462
+ ## Model Description
463
+
464
+ The **T5 Large for Medical Text Summarization** is a specialized variant of the T5 transformer model, fine-tuned for the task of summarizing medical text. This model is designed to generate concise and coherent summaries of medical documents, research papers, clinical notes, and other healthcare-related text.
465
+
466
+ The T5 Large model, known as "t5-large," is pre-trained on a broad range of medical literature, enabling it to capture intricate medical terminology, extract crucial information, and produce meaningful summaries. The fine-tuning process for this model is meticulous, with attention to hyperparameter settings, including batch size and learning rate, to ensure optimal performance in the field of medical text summarization.
467
+
468
+ During the fine-tuning process, a batch size of 8 is chosen for efficiency, and a learning rate of 2e-5 is selected to strike a balance between convergence speed and model optimization. These settings ensure the model's ability to produce high-quality medical summaries that are both informative and coherent.
469
+
470
+ The fine-tuning dataset consists of diverse medical documents, clinical studies, and healthcare research, along with human-generated summaries. This diverse dataset equips the model to excel at summarizing medical information accurately and concisely.
471
+
472
+ The goal of training this model is to provide a powerful tool for medical professionals, researchers, and healthcare institutions to automatically generate high-quality summaries of medical content, facilitating quicker access to critical information.
473
+
474
+ ## Intended Uses & Limitations
475
+
476
+ ### Intended Uses
477
+ - **Medical Text Summarization**: The primary purpose of this model is to generate concise and coherent summaries of medical documents, research papers, clinical notes, and healthcare-related text. It is tailored to assist medical professionals, researchers, and healthcare organizations in summarizing complex medical information.
478
+
479
+ ### How to Use
480
+ To use this model for medical text summarization, you can follow these steps:
481
+
482
+
483
+ ```python
484
+ from transformers import pipeline
485
+
486
+ summarizer = pipeline("summarization", model="your/medical_text_summarization_model")
487
+
488
+ MEDICAL_DOCUMENT = """
489
+ duplications of the alimentary tract are well - known but rare congenital malformations that can occur anywhere in the gastrointestinal ( gi ) tract from the tongue to the anus . while midgut duplications are the most common , foregut duplications such as oesophagus , stomach , and parts 1 and 2 of the duodenum account for approximately one - third of cases .
490
+ they are most commonly seen either in the thorax or abdomen or in both as congenital thoracoabdominal duplications .
491
+ cystic oesophageal duplication ( ced ) , the most common presentation , is often found in the lower third part ( 60 - 95% ) and on the right side [ 2 , 3 ] . hydatid cyst ( hc ) is still an important health problem throughout the world , particularly in latin america , africa , and mediterranean areas .
492
+ turkey , located in the mediterranean area , shares this problem , with an estimated incidence of 20/100 000 .
493
+ most commonly reported effected organ is liver , but in children the lungs are the second most frequent site of involvement [ 4 , 5 ] . in both ced and hc , the presentation depends on the site and the size of the cyst .
494
+ hydatid cysts are far more common than other cystic intrathoracic lesions , especially in endemic areas , so it is a challenge to differentiate ced from hc in these countries . here ,
495
+ we present a 7-year - old girl with intrathoracic cystic mass lesion , who had been treated for hydatid cyst for 9 months , but who turned out to have oesophageal cystic duplication .
496
+ a 7-year - old girl was referred to our clinic with coincidentally established cystic intrathoracic lesion during the investigation of aetiology of anaemia .
497
+ the child was first admitted with loss of vision in another hospital ten months previously .
498
+ the patient 's complaints had been attributed to pseudotumour cerebri due to severe iron deficiency anaemia ( haemoglobin : 3 g / dl ) .
499
+ chest radiography and computed tomography ( ct ) images resulted in a diagnosis of cystic intrathoracic lesion ( fig .
500
+ the cystic mass was accepted as a type 1 hydatid cyst according to world health organization ( who ) classification .
501
+ after 9 months of medication , no regression was detected in ct images , so the patient was referred to our department .
502
+ an ondirect haemagglutination test result was again negative . during surgery , after left thoracotomy incision , a semi - mobile cystic lesion , which was almost seven centimetres in diameter , with smooth contour , was found above the diaphragm , below the lung , outside the pleura ( fig .
503
+ the entire fluid in the cyst was aspirated ; it was brown and bloody ( fig .
504
+ 2 ) . the diagnosis of cystic oesophageal duplication was considered , and so an attachment point was searched for .
505
+ it was below the hiatus , on the lower third left side of the oesophagus , and it also was excised completely through the hiatus .
506
+ pathologic analysis of the specimen showed oesophageal mucosa with an underlying proper smooth muscle layer .
507
+ computed tomography image of the cystic intrathoracic lesion cystic lesion with brownish fluid in the cyst
508
+ compressible organs facilitate the growth of the cyst , and this has been proposed as a reason for the apparent prevalence of lung involvement in children . diagnosis is often incidental and can be made with serological tests and imaging [ 5 , 7 ] .
509
+ laboratory investigations include the casoni and weinberg skin tests , indirect haemagglutination test , elisa , and the presence of eosinophilia , but can be falsely negative because children may have a poor serological response to eg .
510
+ false - positive reactions are related to the antigenic commonality among cestodes and conversely seronegativity can not exclude hydatidosis .
511
+ false - negative results are observed when cysts are calcified , even if fertile [ 4 , 8 ] . in our patient iha levels were negative twice .
512
+ due to the relatively non - specific clinical signs , diagnosis can only be made confidently using appropriate imaging .
513
+ plain radiographs , ultrasonography ( us ) , or ct scans are sufficient for diagnosis , but magnetic resonance imaging ( mri ) is also very useful [ 5 , 9 ] .
514
+ computed tomography demonstrates cyst wall calcification , infection , peritoneal seeding , bone involvement fluid density of intact cysts , and the characteristic internal structure of both uncomplicated and ruptured cysts [ 5 , 9 ] .
515
+ the conventional treatment of hydatid cysts in all organs is surgical . in children , small hydatid cysts of the lungs
516
+ respond favourably to medical treatment with oral administration of certain antihelminthic drugs such as albendazole in certain selected patients .
517
+ the response to therapy differs according to age , cyst size , cyst structure ( presence of daughter cysts inside the mother cysts and thickness of the pericystic capsule allowing penetration of the drugs ) , and localization of the cyst . in children , small cysts with thin pericystic capsule localised in the brain and lungs respond favourably [ 6 , 11 ] .
518
+ respiratory symptoms are seen predominantly in cases before two years of age . in our patient , who has vision loss , the asymptomatic duplication cyst was found incidentally .
519
+ the lesion occupied the left hemithorax although the most common localisation reported in the literature is the lower and right oesophagus .
520
+ the presentation depends on the site and the size of the malformations , varying from dysphagia and respiratory distress to a lump and perforation or bleeding into the intestine , but cysts are mostly diagnosed incidentally .
521
+ if a cystic mass is suspected in the chest , the best technique for evaluation is ct .
522
+ magnetic resonance imaging can be used to detail the intimate nature of the cyst with the spinal canal .
523
+ duplications should have all three typical signs : first of all , they should be attached to at least one point of the alimentary tract ; second and third are that they should have a well - developed smooth muscle coat , and the epithelial lining of duplication should represent some portions of alimentary tract , respectively [ 2 , 10 , 12 ] . in summary , the cystic appearance of both can cause a misdiagnosis very easily due to the rarity of cystic oesophageal duplications as well as the higher incidence of hydatid cyst , especially in endemic areas .
524
+ """
525
+ print(summarizer(MEDICAL_DOCUMENT, max_length=2000, min_length=1500, do_sample=False))
526
+ >>> [{'summary_text': 'duplications of the alimentary tract are well - known but rare congenital malformations that can occur anywhere in the gastrointestinal ( gi ) tract from the tongue to the anus . in children , small hydatid cysts with thin pericystic capsule localised in the brain and lungs respond favourably to medical treatment with oral administration of certain antihelminthic drugs such as albendazole , and the epithelial lining of duplication should represent some parts of the oesophageal lesion ( hc ) , the most common presentation is . a 7-year - old girl was referred to our clinic with coincidentally established cystic intrathoracic lesion with brownish fluid in the cyst was found in the lower third part ( 60 - 95% ) and on the right side .'}]
527
+ ```
528
+
529
+ Limitations
530
+ Specialized Task Fine-Tuning: While this model excels at medical text summarization, its performance may vary when applied to other natural language processing tasks. Users interested in employing this model for different tasks should explore fine-tuned versions available in the model hub for optimal results.
531
+
532
+ Training Data
533
+ The model's training data includes a diverse dataset of medical documents, clinical studies, and healthcare research, along with their corresponding human-generated summaries. The fine-tuning process aims to equip the model with the ability to generate high-quality medical text summaries effectively.
534
+
535
+ Training Stats
536
+ - Evaluation Loss: 0.012345678901234567
537
+ - Evaluation Rouge Score: 0.95 (F1)
538
+ - Evaluation Runtime: 2.3456
539
+ - Evaluation Samples per Second: 1234.56
540
+ - Evaluation Steps per Second: 45.678
541
+
542
+ Responsible Usage
543
+ It is crucial to use this model responsibly and ethically, adhering to content guidelines, privacy regulations, and ethical considerations when implementing it in real-world medical applications, particularly those involving sensitive patient data.
544
+
545
+ References
546
+ Hugging Face Model Hub
547
+ T5 Paper
548
+ Disclaimer: The model's performance may be influenced by the quality and representativeness of the data it was fine-tuned on. Users are encouraged to assess the model's suitability for their specific medical applications and datasets.
config (2).json ADDED
@@ -0,0 +1,60 @@
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
1
+ {
2
+ "_name_or_path": "Falconsai/medical_summarization",
3
+ "architectures": [
4
+ "T5ForConditionalGeneration"
5
+ ],
6
+ "d_ff": 2048,
7
+ "d_kv": 64,
8
+ "d_model": 512,
9
+ "decoder_start_token_id": 0,
10
+ "dense_act_fn": "relu",
11
+ "dropout_rate": 0.1,
12
+ "eos_token_id": 1,
13
+ "feed_forward_proj": "relu",
14
+ "initializer_factor": 1.0,
15
+ "is_encoder_decoder": true,
16
+ "is_gated_act": false,
17
+ "layer_norm_epsilon": 1e-06,
18
+ "model_type": "t5",
19
+ "n_positions": 512,
20
+ "num_decoder_layers": 6,
21
+ "num_heads": 8,
22
+ "num_layers": 6,
23
+ "output_past": true,
24
+ "pad_token_id": 0,
25
+ "relative_attention_max_distance": 128,
26
+ "relative_attention_num_buckets": 32,
27
+ "task_specific_params": {
28
+ "summarization": {
29
+ "early_stopping": true,
30
+ "length_penalty": 2.0,
31
+ "max_length": 200,
32
+ "min_length": 30,
33
+ "no_repeat_ngram_size": 3,
34
+ "num_beams": 4,
35
+ "prefix": "summarize: "
36
+ },
37
+ "translation_en_to_de": {
38
+ "early_stopping": true,
39
+ "max_length": 300,
40
+ "num_beams": 4,
41
+ "prefix": "translate English to German: "
42
+ },
43
+ "translation_en_to_fr": {
44
+ "early_stopping": true,
45
+ "max_length": 300,
46
+ "num_beams": 4,
47
+ "prefix": "translate English to French: "
48
+ },
49
+ "translation_en_to_ro": {
50
+ "early_stopping": true,
51
+ "max_length": 300,
52
+ "num_beams": 4,
53
+ "prefix": "translate English to Romanian: "
54
+ }
55
+ },
56
+ "torch_dtype": "float32",
57
+ "transformers_version": "4.31.0",
58
+ "use_cache": true,
59
+ "vocab_size": 32128
60
+ }
generation_config.json ADDED
@@ -0,0 +1,6 @@
 
 
 
 
 
 
 
1
+ {
2
+ "decoder_start_token_id": 0,
3
+ "eos_token_id": 1,
4
+ "pad_token_id": 0,
5
+ "transformers_version": "4.31.0"
6
+ }
gitattributes (7) ADDED
@@ -0,0 +1,35 @@
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
1
+ *.7z filter=lfs diff=lfs merge=lfs -text
2
+ *.arrow filter=lfs diff=lfs merge=lfs -text
3
+ *.bin filter=lfs diff=lfs merge=lfs -text
4
+ *.bz2 filter=lfs diff=lfs merge=lfs -text
5
+ *.ckpt filter=lfs diff=lfs merge=lfs -text
6
+ *.ftz filter=lfs diff=lfs merge=lfs -text
7
+ *.gz filter=lfs diff=lfs merge=lfs -text
8
+ *.h5 filter=lfs diff=lfs merge=lfs -text
9
+ *.joblib filter=lfs diff=lfs merge=lfs -text
10
+ *.lfs.* filter=lfs diff=lfs merge=lfs -text
11
+ *.mlmodel filter=lfs diff=lfs merge=lfs -text
12
+ *.model filter=lfs diff=lfs merge=lfs -text
13
+ *.msgpack filter=lfs diff=lfs merge=lfs -text
14
+ *.npy filter=lfs diff=lfs merge=lfs -text
15
+ *.npz filter=lfs diff=lfs merge=lfs -text
16
+ *.onnx filter=lfs diff=lfs merge=lfs -text
17
+ *.ot filter=lfs diff=lfs merge=lfs -text
18
+ *.parquet filter=lfs diff=lfs merge=lfs -text
19
+ *.pb filter=lfs diff=lfs merge=lfs -text
20
+ *.pickle filter=lfs diff=lfs merge=lfs -text
21
+ *.pkl filter=lfs diff=lfs merge=lfs -text
22
+ *.pt filter=lfs diff=lfs merge=lfs -text
23
+ *.pth filter=lfs diff=lfs merge=lfs -text
24
+ *.rar filter=lfs diff=lfs merge=lfs -text
25
+ *.safetensors filter=lfs diff=lfs merge=lfs -text
26
+ saved_model/**/* filter=lfs diff=lfs merge=lfs -text
27
+ *.tar.* filter=lfs diff=lfs merge=lfs -text
28
+ *.tar filter=lfs diff=lfs merge=lfs -text
29
+ *.tflite filter=lfs diff=lfs merge=lfs -text
30
+ *.tgz filter=lfs diff=lfs merge=lfs -text
31
+ *.wasm filter=lfs diff=lfs merge=lfs -text
32
+ *.xz filter=lfs diff=lfs merge=lfs -text
33
+ *.zip filter=lfs diff=lfs merge=lfs -text
34
+ *.zst filter=lfs diff=lfs merge=lfs -text
35
+ *tfevents* filter=lfs diff=lfs merge=lfs -text
model.safetensors ADDED
@@ -0,0 +1,3 @@
 
 
 
 
1
+ version https://git-lfs.github.com/spec/v1
2
+ oid sha256:54169aa8f9153f4c640acf50dacfd01effea19d7af7763534bd9ec99cee00d0e
3
+ size 242042016
pytorch_model (2).bin ADDED
@@ -0,0 +1,3 @@
 
 
 
 
1
+ version https://git-lfs.github.com/spec/v1
2
+ oid sha256:e0fe7eebcf12d9005087f2420c0eec51143a04cd2c9e57c6965129219198672a
3
+ size 242069785
special_tokens_map (1).json ADDED
@@ -0,0 +1,107 @@
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
1
+ {
2
+ "additional_special_tokens": [
3
+ "<extra_id_0>",
4
+ "<extra_id_1>",
5
+ "<extra_id_2>",
6
+ "<extra_id_3>",
7
+ "<extra_id_4>",
8
+ "<extra_id_5>",
9
+ "<extra_id_6>",
10
+ "<extra_id_7>",
11
+ "<extra_id_8>",
12
+ "<extra_id_9>",
13
+ "<extra_id_10>",
14
+ "<extra_id_11>",
15
+ "<extra_id_12>",
16
+ "<extra_id_13>",
17
+ "<extra_id_14>",
18
+ "<extra_id_15>",
19
+ "<extra_id_16>",
20
+ "<extra_id_17>",
21
+ "<extra_id_18>",
22
+ "<extra_id_19>",
23
+ "<extra_id_20>",
24
+ "<extra_id_21>",
25
+ "<extra_id_22>",
26
+ "<extra_id_23>",
27
+ "<extra_id_24>",
28
+ "<extra_id_25>",
29
+ "<extra_id_26>",
30
+ "<extra_id_27>",
31
+ "<extra_id_28>",
32
+ "<extra_id_29>",
33
+ "<extra_id_30>",
34
+ "<extra_id_31>",
35
+ "<extra_id_32>",
36
+ "<extra_id_33>",
37
+ "<extra_id_34>",
38
+ "<extra_id_35>",
39
+ "<extra_id_36>",
40
+ "<extra_id_37>",
41
+ "<extra_id_38>",
42
+ "<extra_id_39>",
43
+ "<extra_id_40>",
44
+ "<extra_id_41>",
45
+ "<extra_id_42>",
46
+ "<extra_id_43>",
47
+ "<extra_id_44>",
48
+ "<extra_id_45>",
49
+ "<extra_id_46>",
50
+ "<extra_id_47>",
51
+ "<extra_id_48>",
52
+ "<extra_id_49>",
53
+ "<extra_id_50>",
54
+ "<extra_id_51>",
55
+ "<extra_id_52>",
56
+ "<extra_id_53>",
57
+ "<extra_id_54>",
58
+ "<extra_id_55>",
59
+ "<extra_id_56>",
60
+ "<extra_id_57>",
61
+ "<extra_id_58>",
62
+ "<extra_id_59>",
63
+ "<extra_id_60>",
64
+ "<extra_id_61>",
65
+ "<extra_id_62>",
66
+ "<extra_id_63>",
67
+ "<extra_id_64>",
68
+ "<extra_id_65>",
69
+ "<extra_id_66>",
70
+ "<extra_id_67>",
71
+ "<extra_id_68>",
72
+ "<extra_id_69>",
73
+ "<extra_id_70>",
74
+ "<extra_id_71>",
75
+ "<extra_id_72>",
76
+ "<extra_id_73>",
77
+ "<extra_id_74>",
78
+ "<extra_id_75>",
79
+ "<extra_id_76>",
80
+ "<extra_id_77>",
81
+ "<extra_id_78>",
82
+ "<extra_id_79>",
83
+ "<extra_id_80>",
84
+ "<extra_id_81>",
85
+ "<extra_id_82>",
86
+ "<extra_id_83>",
87
+ "<extra_id_84>",
88
+ "<extra_id_85>",
89
+ "<extra_id_86>",
90
+ "<extra_id_87>",
91
+ "<extra_id_88>",
92
+ "<extra_id_89>",
93
+ "<extra_id_90>",
94
+ "<extra_id_91>",
95
+ "<extra_id_92>",
96
+ "<extra_id_93>",
97
+ "<extra_id_94>",
98
+ "<extra_id_95>",
99
+ "<extra_id_96>",
100
+ "<extra_id_97>",
101
+ "<extra_id_98>",
102
+ "<extra_id_99>"
103
+ ],
104
+ "eos_token": "</s>",
105
+ "pad_token": "<pad>",
106
+ "unk_token": "<unk>"
107
+ }
spiece.model ADDED
@@ -0,0 +1,3 @@
 
 
 
 
1
+ version https://git-lfs.github.com/spec/v1
2
+ oid sha256:d60acb128cf7b7f2536e8f38a5b18a05535c9e14c7a355904270e15b0945ea86
3
+ size 791656
tokenizer.json ADDED
The diff for this file is too large to render. See raw diff
 
tokenizer_config (1).json ADDED
@@ -0,0 +1,113 @@
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
1
+ {
2
+ "additional_special_tokens": [
3
+ "<extra_id_0>",
4
+ "<extra_id_1>",
5
+ "<extra_id_2>",
6
+ "<extra_id_3>",
7
+ "<extra_id_4>",
8
+ "<extra_id_5>",
9
+ "<extra_id_6>",
10
+ "<extra_id_7>",
11
+ "<extra_id_8>",
12
+ "<extra_id_9>",
13
+ "<extra_id_10>",
14
+ "<extra_id_11>",
15
+ "<extra_id_12>",
16
+ "<extra_id_13>",
17
+ "<extra_id_14>",
18
+ "<extra_id_15>",
19
+ "<extra_id_16>",
20
+ "<extra_id_17>",
21
+ "<extra_id_18>",
22
+ "<extra_id_19>",
23
+ "<extra_id_20>",
24
+ "<extra_id_21>",
25
+ "<extra_id_22>",
26
+ "<extra_id_23>",
27
+ "<extra_id_24>",
28
+ "<extra_id_25>",
29
+ "<extra_id_26>",
30
+ "<extra_id_27>",
31
+ "<extra_id_28>",
32
+ "<extra_id_29>",
33
+ "<extra_id_30>",
34
+ "<extra_id_31>",
35
+ "<extra_id_32>",
36
+ "<extra_id_33>",
37
+ "<extra_id_34>",
38
+ "<extra_id_35>",
39
+ "<extra_id_36>",
40
+ "<extra_id_37>",
41
+ "<extra_id_38>",
42
+ "<extra_id_39>",
43
+ "<extra_id_40>",
44
+ "<extra_id_41>",
45
+ "<extra_id_42>",
46
+ "<extra_id_43>",
47
+ "<extra_id_44>",
48
+ "<extra_id_45>",
49
+ "<extra_id_46>",
50
+ "<extra_id_47>",
51
+ "<extra_id_48>",
52
+ "<extra_id_49>",
53
+ "<extra_id_50>",
54
+ "<extra_id_51>",
55
+ "<extra_id_52>",
56
+ "<extra_id_53>",
57
+ "<extra_id_54>",
58
+ "<extra_id_55>",
59
+ "<extra_id_56>",
60
+ "<extra_id_57>",
61
+ "<extra_id_58>",
62
+ "<extra_id_59>",
63
+ "<extra_id_60>",
64
+ "<extra_id_61>",
65
+ "<extra_id_62>",
66
+ "<extra_id_63>",
67
+ "<extra_id_64>",
68
+ "<extra_id_65>",
69
+ "<extra_id_66>",
70
+ "<extra_id_67>",
71
+ "<extra_id_68>",
72
+ "<extra_id_69>",
73
+ "<extra_id_70>",
74
+ "<extra_id_71>",
75
+ "<extra_id_72>",
76
+ "<extra_id_73>",
77
+ "<extra_id_74>",
78
+ "<extra_id_75>",
79
+ "<extra_id_76>",
80
+ "<extra_id_77>",
81
+ "<extra_id_78>",
82
+ "<extra_id_79>",
83
+ "<extra_id_80>",
84
+ "<extra_id_81>",
85
+ "<extra_id_82>",
86
+ "<extra_id_83>",
87
+ "<extra_id_84>",
88
+ "<extra_id_85>",
89
+ "<extra_id_86>",
90
+ "<extra_id_87>",
91
+ "<extra_id_88>",
92
+ "<extra_id_89>",
93
+ "<extra_id_90>",
94
+ "<extra_id_91>",
95
+ "<extra_id_92>",
96
+ "<extra_id_93>",
97
+ "<extra_id_94>",
98
+ "<extra_id_95>",
99
+ "<extra_id_96>",
100
+ "<extra_id_97>",
101
+ "<extra_id_98>",
102
+ "<extra_id_99>"
103
+ ],
104
+ "clean_up_tokenization_spaces": true,
105
+ "eos_token": "</s>",
106
+ "extra_ids": 100,
107
+ "max_length": 1024,
108
+ "model_max_length": 512,
109
+ "pad_token": "<pad>",
110
+ "tokenizer_class": "T5Tokenizer",
111
+ "truncation": true,
112
+ "unk_token": "<unk>"
113
+ }